Chapter
4
Native Device Message Details
4-112
INIT:SEM
Spectrum Emission Mask Initiate
Function
This command performs single Spectrum Emission Mask measurement.
Command
INIT:SEM
Details
When this function is executed, Spectrum Emission Mask measurement
is set to On and the measurement is performed once.
Example of Use
To perform SEM measurement once.
INIT:SEM
INIT:SPUR
Spurious Emission Initiate
Function
This command starts the Spurious Emission measurement.
Command
INIT:SPUR
Details
When this function is executed, the Spurious measurement is set to On
and the measurement starts.
Example of Use
To start the Spurious measurement.
INIT:SPUR