Chapter
4
Native Device Message Details
4-85
4
Native Device Message Details
ERASEWUP
Erase Warm Up Message
Function
This command erases the warm up message displayed after startup.
Command
ERASEWUP
Example of Use
To erase the warm up message.
ERASEWUP
ESE2/ESE2?
End Event Status Enable Command/Query
Function
This command sets the END event status enable register. When the
query command is issued, the value of the END event status enable
register is returned.
Command
ESE2 n
Query
ESE2?
Response
n
Parameter
n
END event status enable register
Value = bit0 + bit1 + bit2 + bit3 + bit4 + bit5 + bit6 +
bit7
bit0: 2
0
= 1 End of sweep
bit1: 2
1
= 2 (Not used)
bit2: 2
2
= 4 (Not used)
bit3: 2
3
= 8 (Not used)
bit4: 2
4
= 16 End of Average
bit5: 2
5
= 32 End of Measure
bit6: 2
6
= 64 End of Max/Min Hold
bit7: 2
7
= 128 (Not used)