Chapter
4
Native Device Message Details
4-54
CALC:MARK:WIDT:TYPE/CALC:MARK:WIDT:TYPE?
Zone Width Type
Function
This command sets the zone width type of the marker.
Command
CALC:MARK:WIDT:TYPE n,zone_type
Query
CALC:MARK:WIDT:TYPE? n
Parameter
n
Marker number
1
Marker 1
2
Marker 2
3
Marker 3
4
Marker 4
5
Marker 5
6
Marker 6
7
Marker 7
8
Marker 8
9
Marker 9
10
Marker 10
When omitted: Marker 1
zone_type
Zone width type
ZONE
Zone marker
SPOT
Spot marker
Details
This command is available when Marker Result is set to Peak.
When Marker Result is set to Integration or Density, Zone width type is
fixed to Zone, Spot cannot be selected.
This command is not available in the following cases:
•
In Time Domain mode
•
During the Spurious Emission measurement
and when Displayed Segment Mode is set to Auto.
•
During the Spectrum Emission Mask measurement.
Example of Use
To set the zone width type of Marker 1 to the zone marker.
CALC:MARK:WIDT:TYPE 1,ZONE
CALC:MARK:WIDT:TYPE? 1
> ZONE