2.14 Spurious Emission Measurement Function
2-817
2
SCPI Device Message Details
[:SENSe]:SPURious:SYNThesis:LPHase ON|OFF|1|0
Low Phase Noise for Spurious Emission Measurement
Function
This command enables/disables Low Phase Noise function during
Spurious Emission measurement.
Command
[:SENSe]:SPURious:SYNThesis:LPHase <switch>
Parameter
<switch>
Low Phase Noise switch
ON|1
Enables Low Phase Noise switch.
OFF|0
Disables Low Phase Noise switch.
Default Off
Details
This function is available when MS2830A-062/066, MS2840A-066/166 is
installed.
This function does not depend on the setting of the Low Phase Noise
switch at System Config. The setting-enabled conditions are as follows:
Function
status
System Config Low
Phase Noise switch
status
Low Phase Noise switch status
Enables the Low Phase Noise function
during Spurious Emission
Disables the Low Phase Noise function
during Spurious Emission
The status at other than spurious emission measurement is reflected by
the System Config switch.
Example of Use
To enable the Low Phase Noise function during Spurious Emission
measurement.
SPUR:SYNT:LPH ON