EasyManua.ls Logo

Anritsu MS2840A - CNVLOSS;CNVLOSS?

Anritsu MS2840A
1452 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter
Native Device Message Details
4-70
CNVLOSS/CNVLOSS?
External Mixer Conversion Loss
Function
This command sets the external mixer's conversion loss.
Command
CNVLOSS power
Query
CNVLOSS?
Response
power
Parameter
power
External mixer's conversion loss
Range 0.00 to 99.99 dB
Resolution 0.01 dB
Suffix code None. Value is returned in dB units.
Default 15.00 dB
Details
This function is available when Option 044/045 is installed for MS2830A.
This function is available when Option 044/046 is installed for MS2840A.
A value is held per one external mixer's band.
Example of Use
To set the external mixer's conversion loss to 10.00 dB.
CNVLOSS 10.00
CNVLOSS?
> 10.00

Table of Contents

Other manuals for Anritsu MS2840A

Related product manuals