Chapter
Native Device Message Details
4-296
SPUR:SWE:TIME/SPUR:SWE:TIME?
Spurious Emission Sweep Time
Function
This command sets the sweep time of each segment.
Command
SPUR:SWE:TIME
seconds_1,seconds_2,seconds_3,seconds_4,seconds_5,second
s_6,seconds_7,seconds_8,seconds_9,seconds_10,seconds_11,
seconds_12,seconds_13,seconds_14,seconds_15,seconds_16,s
econds_17,seconds_18,seconds_19,seconds_20
Query
SPUR:SWE:TIME?
Response
seconds_1,seconds_2,seconds_3,seconds_4,seconds_5,second
s_6,seconds_7,seconds_8,seconds_9,seconds_10,seconds_11,
seconds_12,seconds_13,seconds_14,seconds_15,seconds_16,s
econds_17,seconds_18,seconds_19,seconds_20
Value is returned in s units, without suffix code.
Parameter
seconds_n
Sweep Time of Segment n
Range
[MS269xA]
2 ms to 1000 s
[MS2830A]
1 ms to 1000 s
[MS2840A]
1 ms to 1000 s
Resolution 1 ms(time
1 s)
0.1 s (1 s < time < 10 s)
1 s (10 s
time)
Suffix code NS, US, MS, S
S is used when omitted.
Default Auto value
Example of Use
To set Sweep Time of each segment.
SWE:TIME
0.1,0.1,0.1,0.1,0.1,0.1,0.1,0.1,0.1,0.1,0.2,0.3,0.1,0.1,
0.1,0.1,0.1,0.1,0.1,0.1
SPUR:SWE:TIME?
>
0.100000,0.100000,0.100000,0.100000,0.100000,0.100000,0.
100000,0.100000,0.100000,0.100000,0.100000,0.100000,0.10
0000,0.100000,0.100000,0.100000,0.100000,0.100000,0.1000
00,0.100000