Native Device Message Details
4-258
SEM:SWE:TIME/SEM:SWE:TIME?
Spectrum Emission Mask Sweep Time
Function
This command sets the sweep time in measuring the reference power for
Spectrum Emission Mask measurement.
Command
SEM:SWE:TIME time
Query
SEM:SWE:TIME?
Response
time
Returns a value of s units, without a suffix code.
Parameter
time
Sweep time
Range
[MS269xA]
2 ms to 1000 s
[MS2830A]
1 ms to 1000 s
[MS2840A]
1 ms to 1000 s
Resolution 1 ms (time ≤ 1 s)
0.1 s (1 s < time < 10 s)
1 s (10 s ≤ time)
Suffix code NS,US,MS,S
S is used when omitted.
Details
The setting range of this function is limited depending on the settings of
other parameters. Refer to
MS2690A/MS2691A/MS2692A Signal
Analyzer Operation Manual (Spectrum Analyzer Function Operation)
or
MS2830A/MS2840A Signal Analyzer Operation Manual (Spectrum
Analyzer Function Operation)
for details.
Example of Use
To set the sweep time in measuring the reference power to 20 ms.
SEM:SWE:TIME 20MS
SEM:SWE:TIME?
> 0.020000