2.3 RBW/VBW
2-89
2
SCPI Device Message Details
:CALCulate:BANDwidth|:BWIDth[:RESolution]:AUTO?
Resolution Bandwidth Auto/Manual Query
Function
This command queries the On/Off state of the automatic resolution
bandwidth setting. When Couple Time/Freq. Domain is set to Off, each
parameter is specified in accordance with the already-specified frequency
axis or time axis measurement.
Query
:CALCulate:BANDwidth|:BWIDth[:RESolution]:AUTO?
Response
<switch>
Parameter
MS269xA, MS2830A, MS2840A RBW Mode Normal
<switch>
Automatic setting
0
Automatic resolution bandwidth setting function
is disabled.
1
Automatic resolution bandwidth setting function
is enabled.
MS2830A, MS2840A RBW Mode CISPR
(Only with MS2830A-016/116, MS2840A-016/116)
<switch>
Automatic setting (for CISPR RBW)
0
Automatic setting is set to Off.
1
Automatic setting is set to On.
Example of Use
To query the On/Off state of the automatic resolution bandwidth setting.
CALC:BAND:AUTO?
> 1
Related command
This command has the same function as the following commands.
[:SENSe]:BANDwidth|:BWIDth[:RESolution]:AUTO?
[:SENSe]:ACPower:BANDwidth[:RESolution]:AUTO?
[:SENSe]:CHPower:BANDwidth[:RESolution]:AUTO?
[:SENSe]:OBWidth:BANDwidth[:RESolution]:AUTO?
:CALCulate:ACPower:BANDwidth[:RESolution]:AUTO?
:CALCulate:CHPower:BANDwidth[:RESolution]:AUTO?
:CALCulate:OBWidth:BANDwidth[:RESolution]:AUTO?