2.10 Burst Average Power
2-457
2
SCPI Device Message Details
Details
This function outputs the measurement result of the Burst Average
Power measurement last performed. It outputs the measurement result
in a state that the measurement has already been done, or in a different
style.
You can use READ command if you perform a measurement along with
starting a sweep again.
The return value of this function varies according to the result
mode.(cf.
:SYSTem:RESult:MODE
)
Example of Use
To obtain the measurement result for Burst Average Power
measurement (n = 1 or when omitted, B mode)
FETC:BPOW?
> 0.000010000,-73.930,-73.930,10001,-999.0,-999.0,-999.0,
-999.0,0.099990000,-999.0
:READ:BPOWer|:TXPower[n]?
Burst Average Power Read
Function
This command performs a measurement for Burst Average Power and
outputs the measurement result.
It achieves the same result as when commands are sent in the order of
:INITiate:BPOWer|:TXPower
:FETCh:BPOWer|:TXPower[n]?
:MEASure:BPOWer|:TXPower[n]?
Burst Average Power Measure
Function
This command performs a measurement for Burst Average Power and
outputs the measurement result.
It achieves the same result as when commands are sent in the order of
:CONFigure:BPOWer|:TXPower
:INITiate:BPOWer|:TXPower
:FETCh:BPOWer|:TXPower[n]?