EasyManua.ls Logo

Anritsu MS2840A - [:SENSe]:SEMask:OFFSet[1]:LIST:SWEep[1]:TIME?

Anritsu MS2840A
1452 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
2.13 Spectrum Emission Mask
2-621
2
SCPI Device Message Details
[:SENSe]:SEMask:OFFSet[1]:LIST:SWEep[1]:TIME?
Spectrum Emission Mask Offset Sweep Time Query
Function
This command queries the sweep time for Spectrum Emission Mask
measurement.
Query
[:SENSe]:SEMask:OFFSet[1]:LIST:SWEep[1]:TIME?
Response
<time_n>
Parameter
<time_n>
Sweep time of Offset-n
Range
[MS269xA]
2 ms to 1000 s
[MS2830A]
1 ms to 1000 s
[MS2840A]
1 ms to 1000 s
Resolution 1 ms (time ≤ 1 s)
0.1 s (1 s < time < 10 s)
1 s (10 s ≤ time)
Suffix code None. Value is returned in S units.
Details
The setting range of this function is limited depending on the settings of
other parameters. Refer to the
MS2690A/MS2691A/MS2692A Signal
Analyzer Operation Manual (Spectrum Analyzer Operation)
or
MS2830A/MS2840A Signal Analyzer Operation Manual (Spectrum
Analyzer Operation)
for details.
Example of Use
To query the sweep time of the offset.
SEM:OFFS:LIST:SWE:TIME?
>
0.040000,0.040000,0.040000,0.040000,0.040000,0.040000,0.
040000,0.040000,0.040000,0.040000,0.040000,0.040000

Table of Contents

Other manuals for Anritsu MS2840A

Related product manuals