2.13 Spectrum Emission Mask
2-621
2
SCPI Device Message Details
[:SENSe]:SEMask:OFFSet[1]:LIST:SWEep[1]:TIME?
Spectrum Emission Mask Offset Sweep Time Query
Function
This command queries the sweep time for Spectrum Emission Mask
measurement.
Query
[:SENSe]:SEMask:OFFSet[1]:LIST:SWEep[1]:TIME?
Response
<time_n>
Parameter
<time_n>
Sweep time of Offset-n
Range
[MS269xA]
2 ms to 1000 s
[MS2830A]
1 ms to 1000 s
[MS2840A]
1 ms to 1000 s
Resolution 1 ms (time ≤ 1 s)
0.1 s (1 s < time < 10 s)
1 s (10 s ≤ time)
Suffix code None. Value is returned in S units.
Details
The setting range of this function is limited depending on the settings of
other parameters. Refer to the
MS2690A/MS2691A/MS2692A Signal
Analyzer Operation Manual (Spectrum Analyzer Operation)
or
MS2830A/MS2840A Signal Analyzer Operation Manual (Spectrum
Analyzer Operation)
for details.
Example of Use
To query the sweep time of the offset.
SEM:OFFS:LIST:SWE:TIME?
>
0.040000,0.040000,0.040000,0.040000,0.040000,0.040000,0.
040000,0.040000,0.040000,0.040000,0.040000,0.040000