Chapter 2 SCPI Device Message Details
2-594
[:SENSe]:SEMask: ATTenuation <rel_ampl>|AUTO
Spectrum Emission Mask Attenuator
Function
This command sets the attenuation when measuring the reference power
for Spectrum Emission Mask measurement.
Command
[:SENSe]:SEMask:ATTenuation <rel_ampl>|AUTO
Parameter
<rel_ampl>
Attenuation when measuring the reference
power
Range 0 to 60 dB
Resolution
[MS269xA]
2 dB
[MS2830A]
,
[MS2840A]
2 dB or 10 dB
Refer to “Table 2.4.2-3 Resolution of Input Attenuator” in the
MS2830A/MS2840A Signal Analyzer Operation Manual
(Spectrum Analyzer Function Operation).
Suffix code DB, dB is used when omitted.
AUTO
Sets attenuation automatically.
Details
The setting range of this function is limited depending on the settings of
other parameters. Refer to the
MS2690A/MS2691A/MS2692A Signal
Analyzer Operation Manual (Spectrum Analyzer Function Operation)
or
MS2830A/MS2840A Signal Analyzer Operation Manual (Spectrum
Analyzer Function Operation)
for details.
Example of Use
To set the attenuation when measuring the reference power to 10 dB.
SEM:ATT 10DB