Chapter 2 SCPI Device Message Details
2-76
[:SENSe]:CORRection:IMPedance[:INPut][:MAGNitude] 50|75
Impedance
Function
This command sets the input impedance.
Command
[:SENSe]:CORRection:IMPedance[:INPut][:MAGNitude] 50|75
Parameter
<mode>
Impedance
50
Sets impedance to 50
Ω
. (Default)
75
Sets impedance to 75
Ω
.
Example of Use
To set the impedance to 75
Ω
.
CORR:IMP 75
[:SENSe]:CORRection:IMPedance[:INPut][:MAGNitude]?
Impedance Query
Function
This command queries the input impedance setting.
Query
[:SENSe]:CORRection:IMPedance[:INPut][:MAGNitude]?
Response
<mode>
Parameter
<mode>
Impedance
50
To set the impedance to 50
Ω
.
75
To set the impedance to 75
Ω
.
Example of Use
To query the input impedance setting.
CORR:IMP?
> 75