2.5 Signal Search
2-165
2
SCPI Device Message Details
:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10:MAXimum:NEXT
Next Peak Search
Function
This command searches for the characteristics of the active trace and
moves the marker to a smaller level peak than the present marker level.
Command
:CALCulate:MARKer[n]:MAXimum:NEXT
Parameter
<n>
Marker number
1
Marker 1
2
Marker 2
3
Marker 3
4
Marker 4
5
Marker 5
6
Marker 6
7
Marker 7
8
Marker 8
9
Marker 9
10
Marker 10
When omitted: Marker 1
Details
This command is not available during Spurious Emission measurement
and when Displayed Segment Mode is Auto.
Not available during Spectrum Emission Mask measurement.
Example of Use
To move the marker 1 to the next peak.
CALC:MARK:MAX:NEXT
Related command
This command has the same function as the following command.
:CALCulate:ACPower:MARKer[1]|2|3|4|5|6|7|8|9|10:MAXimum:
NEXT