2.7 Sweep/Trigger/Gate
2-265
2
SCPI Device Message Details
[:SENSe]:SWEep[:TYPE][:AUTO]:RULes:FFT:RWIDth?
Sweep Type Select Rules Real FFT Width Query
Function
This command queries the FFT width that is actually used for the
sweep/FFT switch rule during measurement.
Query
[:SENSe]:SWEep[:TYPE][:AUTO]:RULes:FFT:RWIDth?
Response
<freq>
Parameter
<freq>
FFT width
Range 40 kHz or 2 MHz
Resolution 1 Hz
Suffix code None. Value is returned in Hz units.
Details
This command is not available for MS269x Series.
The specified width is used as the FFT width on a priority basis, but it is
automatically changed by the SPAN/RBW/VBW measurement conditions.
This command queries the FFT width that is actually used.
Example of Use
To query the FFT width that is actually used.
SWE:RUL:FFT:RWID?
> 40000