Chapter 2 SCPI Device Message Details
2-302
[:SENSe]:SWEep:EGATe[:VIEW]:TIME <time>
Gate View Sweep Time
Function
This command sets the sweep time for Gate View.
This parameter and the one for setting the sweep time for time axis
measurement reference the same value.
Command
[:SENSe]:SWEep:EGATe[:VIEW]:TIME <time>
Parameter
<time>
Sweep time
Range 1 µs to 1000 s
Suffix code NS, US, MS, S
S is used when omitted.
Default 100 ms
Details
This command is not available when Gate View is set to Off.
Example of Use
To set the sweep time to 100 ms.
SWE:EGAT:TIME 0.1
[:SENSe]:SWEep:EGATe[:VIEW]:TIME?
Gate View Sweep Time Query
Function
This command queries the sweep time for Gate View.
This parameter and the one for setting the sweep time for time axis
measurement reference the same value.
Query
[:SENSe]:SWEep:EGATe[:VIEW]:TIME?
Response
<time>
Parameter
<time>
Sweep time
Range 1 µs to 1000 s
Suffix code None. Value is returned in s units.