2.13 Spectrum Emission Mask
2-673
2
SCPI Device Message Details
:CALCulate:SEMask:OFFSet[1]:LIST:STOP:RCARrier
<ampl_1>[,<ampl_2>[,<ampl_3>[,<ampl_4>[,<ampl_5>[,<ampl_6>[,<ampl_7
>[,<ampl_8>[,<ampl_9>[,<ampl_10>[,<ampl_11>[,<ampl_12>]]]]]]]]]]]
Spectrum Emission Mask Offset Stop Frequency Limit Level
Function
This command sets the relative level upper limit of the offset stop
frequency for Spectrum Emission Mask measurement.
Command
:CALCulate:SEMask:OFFSet[1]:LIST:STOP:RCARrier <ampl_n>
Parameter
<ampl_n>
Relative level upper limit of Offset-n stop
frequency
Range 200 to +50 dB
Resolution 0.01 dB
Suffix code DB
dB is set when omitted.
Example of Use
To set the absolute level upper limit of the stop frequency of the offset.
CALC:SEM:OFFS:LIST:STOP:RCAR
-30,-30,-30,-30,-30,-30,-30,-30,-30,-30,-30,-30
Related command
This command has the same function as the following command.
[:SENSe]:SEMask:OFFSet[1]:LIST:STOP:RCARrier