EasyManua.ls Logo

Anritsu MS2840A - :CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10:WIDTh <freq>

Anritsu MS2840A
1452 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
2.4 Marker
2-131
2
SCPI Device Message Details
:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10:WIDTh <freq>
Zone Marker Width (by Frequency)
Function
This command sets the zone marker width by the frequency.
Command
:CALCulate:MARKer[n]:WIDTh <freq>
Parameter
<n>
Marker number
1
Marker 1
2
Marker 2
3
Marker 3
4
Marker 4
5
Marker 5
6
Marker 6
7
Marker 7
8
Marker 8
9
Marker 9
10
Marker 10
When omitted:
Marker 1
<freq>
Frequency of zone marker
Range 0.01 Hz to the specified span width
Resolution 0.01 Hz
Suffix code HZ, KHZ, KZ, MHZ, MZ, GHZ, GZ
Hz is used when omitted.
Default Value of 1 Division
Details
This command is not available during Spurious Emission measurement
and when Displayed Segment Mode is Auto.
Not available during Spectrum Emission Mask measurement.
Example of Use
To set the zone marker width of Marker 1 to 1 MHz.
CALC:MARK:WIDT 1MHZ

Table of Contents

Other manuals for Anritsu MS2840A

Related product manuals