Chapter 2 SCPI Device Message Details
2-278
:TRIGger[:SEQuence]:EXTernal[1]:SLOPe POSitive|NEGative
Trigger Slope
Function
This command sets the trigger detection mode (rise/fall).
Command
:TRIGger[:SEQuence]:EXTernal1:SLOPe <mode>
Parameter
<mode>
Trigger detection mode
POSitive
Detects at the rising edge (Default)
NEGative
Detects at the falling edge
Details
This function selects the trigger detection mode when the trigger signal
source is video detection (Video), wideband IF detection (Wide IF Video),
external input (External), SG Marker, or Baseband Interface (BBIF).
Example of Use
To detect at the rising edge.
TRIG:EXT:SLOP POS
Related command
This command has the same function as the following commands.
:TRIGger[:SEQuence]:WIF|:RFBurst:SLOPe
:TRIGger[:SEQuence]:VIDeo:SLOPe