EasyManua.ls Logo

Anritsu MS2840A - 2.17 Configuring Batch Measurement Settings

Anritsu MS2840A
1452 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 2 SCPI Device Message Detail
2-856
2.17 Configuring Batch Measurement Settings
Table 2.17-1 lists device messages for setting the Batch Measurement
function.
Table 2.17-1 Device Messages for Batch Measurement Settings
Function Device Message
Reloading Parameter
List Files
:MMEMory:RELoad:BATCh [<device>]
Adjacent Channel
Power Batch Measure
:MEASure:BATCh:ACP[n]? <filename>[,<device>]
Occupied Bandwidth
Batch Measure
:MEASure:BATCh:OBWidth[n]? <filename>[,<device>]
Spectrum Emission
Mask Batch Measure
:MEASure:BATCh:SEMask[n]? <filename>[,<device>]
Spurious Emission
Batch Measure
:MEASure:BATCh:SPURious[n]? <filename>[,<device>]
Transmit
Intermodulation
Batch Measure
:MEASure:BATCh:IM? <filename>,<spa_freq>[,<sg_freq>[,<device>]]
Measure Power Adjust
:MEASure:POWadj?
<rbw>,<length>,<sg_start_level>,<sg_max_level>,<target>,<range>[,<fr
equency>[,<tracepoint>[,<count>[,<adjust_log>[,<sg_offset_switch>]]]
]]

Table of Contents

Other manuals for Anritsu MS2840A

Related product manuals