SCPI Device Message Details
[:SENSe]:ACPower:FILTer[:RRC]:ALPHa <real>
Adjacent Channel Power Offset Rolloff Ratio
Function
This command sets the rolloff ratio of the filter of the offset channel for
Adjacent Channel Power measurement.
Command
[:SENSe]:ACPower:FILTer[:RRC]:ALPHa <real>
Parameter
<real>
Filter rolloff ratio
Range 0.01 to 1.00
Resolution 0.01
Suffix code None
Details
This function is enabled when the filter type of the offset channel for
ACP measurement(ACP Offset Filter Type) is set to either of the
following:
•
Nyquist
•
Root Nyquist
Example of Use
To set the rolloff ratio of the filter for ACP measurement to 0.22.
ACP:FILT:ALPH 0.22
Related command
This command has the same function as the following command.
:CALCulate:ACPower:FILTer[:RRC]:ALPHa
:CALCulate:ACPower:FILTer[:RRC]:ALPHa <real>
Adjacent Channel Power Offset Rolloff Ratio
Function
This command sets the rolloff ratio of the filter of the offset channel for
Adjacent Channel Power measurement.
Refer to
[:SENSe]:ACPower:FILTer[:RRC]:ALPHa.
Related command
This command has the same function as the following command.
[:SENSe]:ACPower:FILTer[:RRC]:ALPHa