Chapter 2 SCPI Device Message Detail
2-850
:CALCulate:SGControl[:STATe] ON|OFF|1|0
Internal Signal Generator Control
Function
This command enables/disables Internal Signal Generator Control
function.
Command
:CALCulate:SGControl[:STATe] <switch>
Parameter
<switch>
Internal Signal Generator Control function
On/Off
ON|1
Enables Internal Signal Generator Control
function.
OFF|0
Disables Internal Signal Generator Control
function. (Default)
Details
This command is available only for MS2830A.
Available when the following option is installed
・
052/152/352 Internal Signal Generator Control Function
and one of the following is installed at the same time.
・
020/120 3.6GHz Vector Signal Generator
・
021/121 6GHz Vector Signal Generator
・
088/188 3.6GHz Analog Signal Generator
All traces are cleared when switching between On and Off.
This cannot be executed under the following conditions:
・
When Scale Mode is Linear
・
When the Measure function is On
・
In Time Domain
・
When the Trigger function is On
・
When Gate Sweep is On
・
When the Frequency Counter function is On
・
When the level display unit is other than dBm
Example of Use
To enable the Internal Signal Generator Control function.
CALC:SGC ON