Chapter 2 SCPI Device Message Details
2-202
:CALCulate:LLINe[1]|2|3|4|5|6:MARGin <ampl_rel>
Margin Value
Function
This command sets the Offset value of Limit Line.
Command
:CALCulate:LLINe[1]|2|3|4|5|6:MARGin <ampl_rel>
Parameter
<ampl_rel>
Offset value
Range –40.00 to 0.00 (Upper)
0.000 to 40.00 (Lower)
Resolution 0.01 dB
Unit dB
Suffix code DB
dB is used when omitted.
Default 0
Example of Use
To set the Offset value to 10 dB.
CALC:LLIN:MARG 10
:CALCulate:LLINe[1]|2|3|4|5|6:MARGin?
Margin Value Query
Function
This command queries the Offset value of Limit Line.
Query
:CALCulate:LLINe[1]|2|3|4|5|6:MARGin?
Response
<ampl_rel>
Parameter
<ampl_rel>
Offset value
Range –40.00 to 0.00 (Upper)
0.00 to 40.00 (Lower)
Resolution 0.01 dB
Unit dB
Suffix code DB