Chapter
4
Native Device Message Details
4-216
SEM:BAND:AUTO/SEM:BAND:AUTO?
Spectrum Emission Mask Resolution Bandwidth Auto/Manual
Function
This command sets the resolution bandwidth (RBW) of the reference
channel for Spectrum Emission Mask measurement.
Command
SEM:BAND:AUTO switch_com
Query
SEM:BAND:AUTO?
Response
switch_res
Parameter
switch_com
Automatic setting function of resolution
bandwidth (RBW)
ON|1
Sets the automatic setting function to On.
OFF|0
Sets the automatic setting function to Off.
switch_res
Automatic setting function of resolution
bandwidth (RBW)
1
Automatic setting function is On.
0
Automatic setting function is Off.
Details
The setting range of this function is limited depending on the settings of
other parameters. Refer to
MS2690A/MS2691A/MS2692A Signal
Analyzer Operation Manual (Spectrum Analyzer Function Operation)
or
MS2830A/MS2840A Signal Analyzer Operation Manual (Spectrum
Analyzer Function Operation)
for details.
Example of Use
To set the resolution bandwidth of the reference channel automatically.
SEM:BAND:AUTO ON
SEM:BAND:AUTO?
> 1