Chapter 2 SCPI Device Message Details
2-46
[:SENSe]:FREQuency:STOP <freq>
Stop Frequency
This command sets the stop frequency.
Command
[:SENSe]:FREQuency:STOP <freq>
Parameter
<freq>
Stop frequency
Range
[MS269xA]
–99.9997 MHz to 6.05 GHz (MS2690A)
–99.9997 MHz to 13.6 GHz (MS2691A)
–99.9997 MHz to 26.6 GHz (MS2692A)
[MS2830A]
–99.9997 MHz to 3.7 GHz (Option 040)
–99.9997 MHz to 6.1 GHz (Option 041)
–99.9997 MHz to 13.6 GHz (Option 043)
–99.9997 MHz to 26.6 GHz (Option 044)
–99.9997 MHz to 43.1 GHz (Option 045)
[MS2840A]
–99.9997 MHz to 3.7 GHz (Option 040)
–99.9997 MHz to 6.1 GHz (Option 041)
–99.9997 MHz to 27 GHz (Option 044)
–99.9997 MHz to 45 GHz (Option 046)
Resolution 1 Hz
Suffix code HZ, KHZ, KZ, MHZ, MZ, GHZ, GZ
Hz is used when omitted.
Default
[MS269xA]
6.00 GHz (MS2690A)
13.50 GHz (MS2691A)
26.50 GHz (MS2692A)
[MS2830A]
3.6 GHz (Option 040)
6.0 GHz (Option 041)
13.5 GHz (Option 043)
26.5 GHz (Option 044)
43 GHz (Option 045)
[MS2840A]
3.6 GHz (Option 040)
6.0 GHz (Option 041)
26.5 GHz (Option 044)
44.5 GHz (Option 046)
Details
This command is not available during the Spectrum Emission Mask
measurement or the Spurious Emission measurement.
When Internal Signal Generator Control On/Off is On, the stop frequency
range depends on the frequency range of Signal Generator.