Chapter
Native Device Message Details
4-68
CMK?
Marker Position Query
Function
This command queries the marker position using the displayed point
from the screen left edge.
Query
CMK? marker
Response
point
Parameter
point
Marker position (Number of displayed points
from the screen left edge)
Range 0 to 10, 20, 40, 50, 100, 200, 250, 400, 500,1000,
2000, 5000, 10000, 30000
(The upper limit value varies according to the
number of trace displayed points.)
Resolution 1
marker
Marker type
1
Marker1
2
Marker2
3
Marker3
4
Marker4
5
Marker5
6
Marker6
7
Marker7
8
Marker8
9
Marker9
10
Marker10
When omitted Active marker
Details
*** is returned when the Marker Mode is Off.
This command is not available during the Spurious Emission
measurement and when Displayed Segment Mode is set to Auto.
Example of Use
To query the active marker position.
CMK?
> 123