Chapter 2 SCPI Device Message Detail
2-944
:MEASure:POWadj?
<rbw>,<length>,<sg_start_level>,<sg_max_level>,<target>,<range>[,<freque
ncy>[,<tracepoint>[,<count>[,<adjust_log>[,<sg_offset_switch>]]]]]
Measure Power Adjust
Function
This command executes power adjustment and reads the result.
Power adjustment combines the SG option and power measurement
functions to set the output level of the DUT, such as an amplifier, to the
specified level.
Query
:MEASure:POWadj?
<rbw>,<length>,<sg_start_level>,<sg_max_level>,<target>,
<range>[,<frequency>[,<tracepoint>[,<count>[,<adjust_log
>[,<sg_offset_switch>]]]]]
Response
<judge>,<sa_input>,<sg_output>,<count_res>,<time>,<sa_in
put_log_n>,<sg_output_log_n>,<count_log_n>
Parameter
<rbw>
Analysis bandwidth
[MS269xA]
Range/Resolution
30 Hz to 31.25 MHz
Either of the values, 30 Hz to 3 MHz (1-3
sequence), 50 kHz, 5 MHz, 10 MHz, 20 MHz,
and 31.25 MHz can be used.
Set the specified waveform bandwidth or more.
Recommended to set at 31.25 MHz
[MS2830A]
[MS2840A]
Range/Resolution
30 Hz to 31.25 MHz
Either of the values, 30 Hz to 3 MHz (1-3
sequence), 500 Hz, 50 kHz, 2 MHz, 5 MHz, 10
MHz, 20 MHz, and 31.25 MHz can be used.
Set the specified waveform bandwidth or more.
Recommended to set at 31.25 MHz
[Common]
Suffix code HZ, KHZ, KZ, MHZ, MZ, GHZ, GZ
Hz is used when omitted
<length>
Analysis length
Range
1 µs to 1000 s
Suffix code
NS,US,MS,S
S is used when omitted
µs (resolution)