2.17 Configuring Batch Measurement Settings
2-945
2
SCPI Device Message Details
<sg_start_level>
Set the SG output level set at the start.
When inserting a DUT, specify the DUT input
level and RF Input so the maximum input level
is not exceeded.
Range
[MS269xA]
140 dBm to
5 dBm
[MS2830A], [MS2840A]
40.00 dBm to +20.00 dBm (>25 MHz)
40.00 dBm to +2.00 dBm (
25 MHz)
136.00 dBm to +15.00 dBm (
>
25 MHz)
(Option 022/122)
136.00 dBm to
3.00 dBm (
25 MHz)
(Option 022/122)
Resolution 0.01 dBm
<sg_max_level>
Set the SG maximum output level.
When inserting a DUT, specify the DUT input
level and RF Input so the maximum input level
is not exceeded.
Range
[MS269xA]
140 dBm to 5 dBm
[MS2830A], [MS2840A]
40.00 dBm to +20.00 dBm (
>
25 MHz)
40.00 dBm to +2.00 dBm (
25 MHz)
136.00 dBm to +15.00 dBm (
>
25 MHz)
(Option 022/122)
136.00 dBm to 3.00 dBm (
25 MHz)
(Option 022/122)
Resolution 0.01 dBm
Suffix code
DBM,DM
<target>
Target level when executing power adjustment
Range
150 dBm to 30 dBm (Pre-Amp Off)
150 dBm to 10 dBm (Pre-Amp On)
When adding the reference offset, the added
value is the setting range.
Resolution 0.01 dBm
Suffix code
DBM,DM
<range>
Power adjustment range
Power adjusted is evaluated as PASS when
adjustment is within the specified range
Range
0 to 20 dB
Resolution 0.01 dB
<frequency>
Power adjustment execution frequency
Range
[MS269xA]
125 MHz to 6 GHz
[MS2830A]
250 kHz to 3.6 GHz (Option 020/120)
250 kHz to 6 GHz (Option 041/043, and 021/121)
[MS2840A]
250 kHz to 3.6 GHz (Option 020/120)