Chapter 2 SCPI Device Message Detail
2-946
250 kHz to 6 GHz (Option 041, and 021/121)
Resolution 1 Hz
Suffix code HZ, KHZ, KZ, MHZ, MZ, GHZ, GZ
Hz is used when omitted.
MAXimum, MINimum, and DEFault cannot be
used. The currently set frequency cannot be used
when omitted.
<tracepoint>
trace point
11
11 point
21
21 point
41
41 point
51
51 point
101
101 point
201
201 point
251
251 point
401
401 point
501
501 point
1001
1001 point
2001
2001 point
5001
5001 point
10001
10001 point
30001
30001 point (MS269xA Only)
Recommended to specify Trace Point to 1001.
<count>
Power adjustment execution count
Range
1 to 10
Resolution 1
Suffix code
None. If the value is omitted, power adjustment
is executed 5 times.
<adjust_log>
Specifies whether or not to output Log at each
power adjustment.
ON|1
Output log
OFF|0
Do not output log
<sg_offset_switch>
Specifies whether or not to output Log at each
power adjustment.
ON|1
Return level offset setting
OFF|0
Do not return the level offset setting values
<judge>
Evaluates power adjustment
0 is returned at PASS and 1 is returned at FAIL.
999.0 is returned if there is no measurement.
<sa_input>
DUT output level
Suffix code
None, dBm (unit)
0.01 dB (resolution)
999.0 is returned if there is no measurement.
<sg_output>
Output level of SG Output of this equipment
Suffix code
None, dBm (unit)
0.01 dB (resolution)
999.0 is returned if there is no measurement.
<count_res>
Power adjustment execution count