Chapter
Native Device Message Details
4-110
GLN/GLN?
Gate Length
Function
This command sets the gate time length.
Command
GLN time
Query
GLN?
Response
time
No suffix code. Value is returned in
µ
s units.
Parameter
time
Delay time until starting gate
Range 50
µ
s to 1 s
Resolution 20 ns
Suffix code NS, US, MS, S
MS is used when omitted.
Example of Use
To set the gate time length to 20 ms.
GLN 20
GLN 20000US
GLN 20MS
GLN 0.02S
GLN?
> 20000