EasyManua.ls Logo

Anritsu MS2840A - :TRIGger:SEMask[:SEQuence]:SOURce EXTernal[1]|IMMediate|WIF|RFBurst|VIDeo|SG|BBIF|FRAMe

Anritsu MS2840A
1452 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 2 SCPI Device Message Details
2-698
:TRIGger:SEMask[:SEQuence]:SOURce
EXTernal[1]|IMMediate|WIF|RFBurst|VIDeo|SG|BBIF|FRAMe
Trigger Source
Function
This command selects the trigger source.
Command
:TRIGger:SEMask[:SEQuence]:SOURce <source>
Parameter
<source>
Trigger signal source
[MS269xA]
EXTernal[1]
External input (External)
IMMediate
Free run
WIF|RFBurst
Wideband IF detection (Wide IF Video)
VIDeo
Video detection(Video)
SG
SG marker
BBIF
Baseband Interface (BBIF)
[MS2830A], [MS2840A]
EXTernal[1]
External input (External)
IMMediate
Free run
WIF|RFBurst
Wideband IF detection (Wide IF Video)
VIDeo
Video detection(Video)
SG
SG marker
FRAMe
Frame period trigger
Details
[MS269xA]
SG marker trigger can be selected only when the Option 020/120 Vector
Signal Generator is installed.
BBIF trigger cannot be selected only when the Option 040/140 Baseband
Interface Unit is not installed or the software package is Ver.6.00.00 or
later.
[MS2830A], [MS2840A]
SG marker trigger can be selected only when the Option 020/120/021/121
Vector Signal Generator is installed.
Example of Use
To set the trigger signal source to Video trigger.
TRIG:SEM:SOUR VID