2.9 ACP
2-369
2
SCPI Device Message Details
:CALCulate:ACPower:CARRier[1]:RCFRequency <freq>
Adjacent Channel Power In Band Center
Function
This command sets the In Band center frequency for Adjacent Channel
Power measurement.
Refer to
[:SENSe]:ACPower:CARRier[1]:RCFRequency.
Related command
This command has the same function as the following command.
[:SENSe]:ACPower:CARRier[1]:RCFRequency
[:SENSe]:ACPower:CARRier[1]:RCFRequency?
Adjacent Channel Power In Band Center Query
Function
This command queries the In Band center frequency for Adjacent
Channel Power measurement.
Query
[:SENSe]:ACPower:CARRier[1]:RCFRequency?
Response
<freq>
Parameter
<freq>
In Band center frequency
Range (Start Frequency) to (Stop Frequency)
Resolution 1 Hz
Suffix code None. Value is returned in Hz unit.
Example of Use
To query the In Band center frequency.
ACP:CARR:RCFR?
> 3000000000
Related command
This command has the same function as the following command.
:CALCulate:ACPower:CARRier[1]:RCFRequency?