Chapter 2 SCPI Device Message Details
2-274
Related command
This command has the same function as the following commands.
:TRIGger:ACPower[:SEQuence]:SOURce?
:TRIGger:CHPower[:SEQuence]:SOURce?
:TRIGger:OBWidth[:SEQuence]:SOURce?
:TRIGger:SEMask[:SEQuence]:SOURce?
:TRIGger:BPOWer|:TXPower[:SEQuence]:SOURce?
:TRIGger[:SEQuence]:VIDeo:LEVel[:LOGarithmic] <level>
Log Scale Video Trigger Level
Function
This command sets the threshold value of the level where trigger sweep
starts.
Command
:TRIGger[:SEQuence]:VIDeo:LEVel[:LOGarithmic] <level>
Parameter
<level>
Threshold value of level where trigger sweep
starts.
Range –150 to +50 dBm
Resolution 1 dB
Suffix code DBM, DM
Default –40 dBm
Example of Use
To set the threshold value of the level where trigger sweep starts to –10
dBm.
TRIG:VID:LEV –10DBM