Chapter
4
Native Device Message Details
4-238
SEM:OFFS:LIST:FREQ:STOP/SEM:OFFS:LIST:FREQ:STOP?
Spectrum Emission Mask Offset Stop Frequency
Function
This command sets the stop frequency of the offset for Spectrum
Emission Mask measurement.
Command
SEM:OFFS:LIST:FREQ:STOP
freq_1[,freq_2[,freq_3[,freq_4[,freq_5[,freq_6[,freq_7[,
freq_8[,freq_9[,freq_10[,freq_11[,freq_12]]]]]]]]]]]
Query
SEM:OFFS:LIST:FREQ:STOP?
Response
freq_1,freq_2,freq_3,freq_4,freq_5,freq_6,freq_7,freq_8,
freq_9,freq_10,freq_11,freq_12
No suffix code. Value is returned in Hz units.
Parameter
freq_n
Stop frequency of Offset-n
Range 300 to 500000000 Hz
Resolution 2 Hz
Suffix code HZ,KHZ,KZ,MHZ,MZ,GHZ,GZ
Hz is used when omitted.
Example of Use
To set the stop frequency of the offset.
SEM:OFFS:LIST:FREQ:STOP
6MHZ,6MHZ,6MHZ,6MHZ,6MHZ,6MHZ,6MHZ,6MHZ,6MHZ,6MHZ,6MHZ,6
MHZ
SEM:OFFS:LIST:FREQ:STOP?
>
6000000,6000000,6000000,6000000,6000000,6000000,6000000,
6000000,6000000,6000000,6000000,6000000