2.7 Sweep/Trigger/Gate
2-253
2
SCPI Device Message Details
:INITiate:SWP?
Single Sweep Query
Function
This command queries the sweep status (sweep done/during sweep).
Query
:INITiate:SWP?
Response
<status>
Parameter
<status>
Sweep status
0
Sweep done
1
During sweep
Example of Use
To query the sweep status.
INIT:SWP?
> 0
[:SENSe]:SWEep:TIME <time>
Sweep Time
Function
This command sets the sweep time. Each parameter is specified in
accordance with the already-specified frequency axis or time axis
measurement.
Command
[:SENSe]:SWEep:TIME <time>
Parameter
<time>
Sweep Time
Range
[MS269xA]
2 ms to 1000 s (in frequency axis measurement)
1 µs to 1000 s (in time axis measurement)
[MS2830A]
1 ms to 1000 s (in frequency axis measurement)
1 µs to 1000 s (in time axis measurement)
[MS2840A]
1 ms to 1000 s (in frequency axis measurement)
1 µs to 1000 s (in time axis measurement)
Suffix code None
Value is returned in s units.
Default
[MS269xA]
10 ms (MS2690A)
135 ms (MS2691A)
265 ms (MS2692A)
[MS2830A]
1 ms (Option 040)