EasyManua.ls Logo

Anritsu MS2840A - [:SENSe]:POWer[:RF]:PADJust?

Anritsu MS2840A
1452 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 2 SCPI Device Message Detail
2-958
[:SENSe]:POWer[:RF]:PADJust?
Pre-selector Tune Query
Function
This command queries the Preselector peaking bias value.
Query
[:SENSe]:POWer[:RF]:PADJust?
Response
<freq>
Parameter
<freq>
Peaking bias value
Range 128 to 127
Resolution 1
Details
[MS269xA]
Automatic setting of the peaking bias value cannot be used under the
following conditions:
When using the MS2690A:
When not using the preselector band:
The frequency which switched to the preselector band can be set in
Frequency Band Mode.
[MS2830A],
[MS2840A]
Automatic setting of the peaking bias value cannot be used under the
following conditions:
MS2830A-040/041/043, MS2840A-040/041:
When not using the preselector band:
The frequency which switched to the preselector band can be set in
Frequency Band Mode.
Example of Use
To query the peaking bias value.
POW:PADJ?
> 100

Table of Contents

Other manuals for Anritsu MS2840A