Chapter 2 SCPI Device Message Details
2-268
Related command
This command has the same function as the following commands.
[:SENSe]:ACPower:DETector[:FUNCtion]
[:SENSe]:CHPower:DETector[:FUNCtion]
[:SENSe]:OBWidth:DETector[:FUNCtion]
:CALCulate:DETector[:FUNCtion]
:CALCulate:ACPower:DETector[:FUNCtion]
:CALCulate:CHPower:DETector[:FUNCtion]
:CALCulate:OBWidth:DETector[:FUNCtion]
[:SENSe]:DETector[:FUNCtion]?
Detection Mode Query
Function
This command selects the detection mode of the waveform pattern. When
Couple Time/Freq. Domain is set to Off, each parameter is specified in
accordance with the already-specified frequency axis or time axis
measurement.
Query
[:SENSe]:DETector[:FUNCtion]?
Response
<det>
Parameter
<det>
Detection mode selection
NORM
Simultaneous positive and negative peak
detection (Default)
POS
Positive peak detection
NEG
Negative peak detection
SAMP
Sample detection
RMS
RMS detection
QPE
QP detection
CAV
CISPR Average detection
CRMS
RMS Average detection
Details
The set detection mode applies to all traces.