Chapter 2 SCPI Device Message Detail
2-956
:CALibration:YTF?
Pre-selector Auto Tune Query
Function
This command queries the pre-selector settings.
Query
:CALibration:YTF?
Response
<result>
Parameter
<result>
Normal/Error
0
Finished normal
1
Error
Details
[MS269xA], [MS2830A], [MS2840A]
This command is not available during the Spurious Emission
measurement and when Displayed Segment Mode is set to Auto.
[MS2830A]
This command is not available for MS2830A-040/041/043.
[MS2840A]
This command is not available for MS2840A-040/041.
Example of Use
To query the pre-selector settings.
CAL:YTF?
> 0