Chapter
Native Device Message Details
4-207
4
Native Device Message Details
RMKL?
Reference Marker Level Query
Function
This command queries the level data of the marker point set by Relative
To of the active marker.
Query
RMKL?
Response
level
Level of the marker point set by Relative To
No suffix code, in dB units, resolution: 0.001 dB
(When marker level display units are dB-system units)
No suffix code, in
µ
V units, resolution: 0.01 pV
(When marker level display units are V-system units)
No suffix code, in
µ
W units, resolution: 0.01 yW
(When marker level display units are W-system units)
Details
Returns *** when Marker Mode is set to other than Delta.
This command is not available during the Spurious Emission
measurement and when Displayed Segment Mode is set to Auto.
Example of Use
To query the level of the marker point set by Relative To of the active
marker.(1.234 dBm).
RMKL?
> 1.234