[:SENSe]:SEMask:OFFSet[1]:LIST:BANDwidth[1]:INTegration:AUTO?
Spectrum Emission Mask Offset Integrate Bandwidth Auto/Manual Query
Function
This command queries the On/Off state of the automatic Integrate BW
setting function for Spectrum Emission Mask measurement.
Query
[:SENSe]:SEMask:OFFSet[1]:LIST:BANDwidth[1]:INTegration:
AUTO?
Response
<switch_n>
Parameter
<switch_n>
Automatic Integrate BW setting function On/Off
state for Offset-n
1
Automatic Integrate BW setting function is
enabled.
0
Automatic Integrate BW setting function is
disabled.
Details
The setting range of this function is limited depending on the setting of
RBW of the target offset. Refer to the
MS2690A/MS2691A/MS2692A
Signal Analyzer Operation Manual (Spectrum Analyzer Function
Operation)
or
MS2830A/MS2840A Signal Analyzer Operation Manual
(Spectrum Analyzer Function Operation)
for details.
This command is not available if Detection of the target offset is set to
any of the following:
Pos&Neg
Positive
Negative
Sample
Example of Use
To query the automatic Integrate BW setting function On/Off state for
the offset
SEM:OFFS:LIST:BAND:INT:AUTO?
> 1,1,1,1,1,1,1,1,1,1,1,1