Chapter
Native Device Message Details
4-153
4
Native Device Message Details
MEAS:POW?
Measure Power Adjust
Function
This command executes power adjustment and reads the result.
Power adjustment combines the SG option and power measurement
functions to set the output level of the DUT, such as an amplifier, to the
specified level.
Query
MEAS:POW?
rbw,length,sg_start_level,sg_max_level,target,range[,fre
quency[,tracepoint[,count[,adjust_log[,sg_offset_switch]
]]]]
Response
judge,sa_input,sg_output,count_res,time,sa_input_log_n,s
g_output_log_n,count_log_n
Parameter
rbw
Analysis bandwidth
[MS269xA]
Range/Resolution 30 Hz to 31.25 MHz
Either of the values, 30 Hz to 3 MHz (1-3
sequence), 50 kHz, 5 MHz, 10 MHz, 20 MHz,
and 31.25 MHz can be used.
Set the specified waveform bandwidth or more.
Recommended to set at 31.25 MHz
[MS2830A], [MS2840A]
Range/Resolution 30 Hz to 31.25 MHz
Either of the values, 30 Hz to 3 MHz (1-3
sequence), 500 Hz, 50 kHz, 2 MHz, 5 MHz, 10
MHz, 20 MHz, and 31.25 MHz can be used.
Set the specified waveform bandwidth or more.
Recommended to set at 31.25 MHz
[Common]
Suffix code HZ, KHZ, KZ, MHZ, MZ, GHZ, GZ
Hz is used when omitted.
length
Analysis length
Range 1 µs to 1000 s
Suffix code NS,US,MS,S
S is used when omitted
µs (resolution)
sg_start_level
Set the SG output level set at the start.
When inserting a DUT, specify the DUT input
level and RF Input so the maximum input level

Table of Contents

Other manuals for Anritsu MS2840A

Related product manuals