SCPI Device Message Details
[:SENSe]:POWer[:RF]:ATTenuation <rel_ampl>
RF Attenuator
Function
This command sets the attenuation.
Command
[:SENSe]:POWer[:RF]:ATTenuation <rel_ampl>
Parameter
<rel_ampl>
Attenuation
Range 0 to 60 dB
Resolution
[MS269xA]
2 dB
[MS2830A]
,
[MS2840A]
2 dB or 10 dB
Refer to “Table 2.4.2-3 Resolution of Input Attenuator” in the
MS2830A/MS2840A Signal Analyzer Operation Manual
(Spectrum Analyzer Function Operation)
.
Suffix code DB, dB is used even when omitted.
Default value 10 dB
Details
This command is not available in the following situations:
•
When Spurious Emission is On AND when Gate View is Off.
Example of Use
To set the attenuation to 10 dB.
POW:ATT 10