Chapter 2 SCPI Device Message Details
2-666
:CALCulate:SEMask:OFFSet[1]:LIST:STOP:ABSolute[1]|2?
Spectrum Emission Mask Offset Stop Frequency Absolute Limit Level Query
Function
This command queries the absolute level upper limit of the offset stop
frequency for Spectrum Emission Mask measurement.
Query
:CALCulate:SEMask:OFFSet[1]:LIST:STOP:ABSolute[1]|2?
Response
<ampl_n>
Parameter
<ampl_n>
Absolute level upper limit of Offset-n stop
frequency
Range –200 to +50 dBm
Resolution 0.01 dB
Suffix code None. Value is returned in dBm units.
Details
ABSolute1 sets the absolute level upper limit 1, and ABSolute2 sets the
absolute level upper limit 2.
Example of Use
To set the absolute level upper limit 2 of the stop frequency of the offset.
CALC:SEM:OFFS:LIST:STOP:ABS2?
>
-10.00,-10.00,-10.00,-5.00,-5.00,-10.00,-10.00,-10.00,-1
0.00,-5.00,-5.00,-10.00
Related command
This command has the same function as the following command.
[:SENSe]:SEMask:OFFSet[1]:LIST:STOP:ABSolute[1]|2?