Chapter 2 SCPI Device Message Detail
2-974
:STATus:QUEStionable:MEASure:NTRansition <integer>
Questionable Measure Register Negative Transition
Function
This command sets the transition filter (negative direction change) of the
QUEStionable Measure register.
Command
:STATus:QUEStionable:MEASure:NTRansition <integer>
Parameter
<integer>
Total byte of transition filter (negative direction
change)
Resolution 1
Range 0 to 65535
Details
This command is available only in the SCPI mode.
Example of Use
To set the transition filter (negative direction change) of the
QUEStionable Measure register to 16.
STAT:QUES:MEAS:NTR 16
:STATus:QUEStionable:MEASure:NTRansition?
Questionable Measure Register Negative Transition Query
Function
This command queries the transition filter (negative direction change) of
the QUEStionable Measure register.
Query
:STATus:QUEStionable:MEASure:NTRansition?
Response
<integer>
Parameter
<integer>
Total byte of transition filter (negative direction
change)
Resolution 1
Range 0 to 65535