EasyManua.ls Logo

Anritsu MS2840A - LOGSCALEDIV;LOGSCALEDIV?

Anritsu MS2840A
1452 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter
Native Device Message Details
4-138
LOGSCALEDIV/LOGSCALEDIV?
Log Scale Range
Function
This command sets the Y-axis scale magnification for the log scale
display.
Command
LOGSCALEDIV scale
Query
LOGSCALEDIV?
Response
scale
No suffix code. Value is returned in dB units.
Parameter
scale
Y-axis scale magnification
0.1
0.1 dB/div
0.2
0.2 dB/div
0.5
0.5 dB/div
1
1 dB/div
2
2 dB/div
5
5 dB/div
10
10 dB/div
20
20 dB/div
Example of Use
To set the scale magnification to 0.5 dB/div.
LOGSCALEDIV 0.5
LOGSCALEDIV?
> 0.5