2
SCPI Device Message Details
:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10:ACTive OFF|ON|0|1
Active Marker
Function
This command specifies the active marker.
Command
:CALCulate:MARKer[n]:ACTive <switch>
Parameter
<n>
Marker number
1
Marker 1
2
Marker 2
3
Marker 3
4
Marker 4
5
Marker 5
6
Marker 6
7
Marker 7
8
Marker 8
9
Marker 9
10
Marker 10
When omitted:
Marker 1
<switch>
Active marker
ON|1
Sets the marker n to On.
OFF|0
Sets the marker n to Off.
Details
Multiple markers can not be set to active.
This command is not available during Spurious Emission measurement
and when Displayed Segment Mode is Auto.
Not available during Spectrum Emission Mask measurement.
Example of Use
To set the marker 1 to active.
CALC:MARK:ACT ON