Chapter 2 SCPI Device Message Detail
2-796
:CONFigure:SPURious
Spurious Emission Configure
Function
This command sets the Spurious Emission measurement to On.
Command
:CONFigure:SPURious
Details
No measurement is performed.
When Spurious Emission measurement is set to On, the active trace is
set to A.
Example of Use
To set the Spurious measurement to On.
CONF:SPUR
:INITiate:SPURious
Spurious Emission Initiate
Function
This command starts the Spurious Emission measurement.
Command
:INITiate:SPURious
Details
When this function is executed, the Spurious measurement is set to On
and the measurement starts.
To read out the measurement results after executing this command,
perform the synchronization control using the “*WAI” command.
Example of Use
To start the Spurious measurement.
INIT:SPUR