EasyManua.ls Logo

Anritsu MS2840A - CALC:MARK:PEAK:SORT:COUN;CALC:MARK:PEAK:SORT:COUN?

Anritsu MS2840A
1452 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter
Native Device Message Details
4-45
4
Native Device Message Details
CALC:MARK:PEAK:SORT:COUN/CALC:MARK:PEAK:SORT:COUN?
Search Peaks Number
Function
This command sets the search number when Search Peaks Sort Y/X is
executed.
Command
CALC:MARK:PEAK:SORT:COUN integer
Query
CALC:MARK:PEAK:SORT:COUN?
Response
integer
Parameter
integer
Search number
Range 1 to 10
Resolution 1
Default 10
Details
This command is not available in the following cases:
During the Spurious Emission measurement and when Displayed
Segment Mode is set to Auto.
During the Spectrum Emission Mask measurement.
Example of Use
To set the search number to 6.
CALC:MARK:PEAK:SORT:COUN 6
CALC:MARK:PEAK:SORT:COUN?
> 6