2.7 Sweep/Trigger/Gate
2-301
2
SCPI Device Message Details
[:SENSe]:SWEep:EGATe:VIEW[:STATe] ON|OFF|1|0
Gate View
Function
This command sets whether to display the gate view.
Command
[:SENSe]:SWEep:EGATe:VIEW[:STATe] <switch>
Parameter
<switch>
Displays or hides Gate View.
OFF|0
Hides Gate View.
ON|1
Displays Gate View.
Details
This command is not available during time domain measurement.
This command is not available when Gate Sweep is set to Off.
This command is not available when the N dB Bandwidth measurement
function is On.
This command is not available when the Math function is other than Off.
Example of Use
To display Gate View.
SWE:EGAT:VIEW ON
[:SENSe]:SWEep:EGATe:VIEW[:STATe]?
Gate View Query
Function
This command queries whether to display the gate view.
Query
[:SENSe]:SWEep:EGATe:VIEW[:STATe]?
Response
<switch>
Parameter
<switch>
Display/hide gate view
0
Gate View is hidden.
1
Gate View is displayed.
Example of Use
To query whether Gate View is displayed.
SWE:EGAT:VIEW?