EasyManua.ls Logo

Anritsu MS2840A - SPUR:PEAK:THR;SPUR:PEAK:THR?

Anritsu MS2840A
1452 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter
Native Device Message Details
4-289
4
Native Device Message Details
SPUR:PEAK:THR/SPUR:PEAK:THR?
Spurious Emission Search Threshold Level
Function
This command sets the spurious search threshold when Result Type is
Peak.
Command
SPUR:PEAK:THR
real_1,real_2,real_3,real_4,real_5,real_6,real_7,real_8,
real_9,real_10,real_11,real_12,real_13,real_14,real_15,r
eal_16,real_17,real_18,real_19,real_20
Query
SPUR:PEAK:THR?
Response
real_1,real_2,real_3,real_4,real_5,real_6,real_7,real_8,
real_9,real_10,real_11,real_12,real_13,real_14,real_15,r
eal_16,real_17,real_18,real_19,real_20
Value is returned in dBm units, without suffix code.
Parameter
real_n
Spurious search threshold value of Segment n
Range 200 dBm to 50 dBm
Resolution 0.01 dB
Suffix code DBM,DM
Example of Use
To set the spurious search threshold.
SPUR:PEAK:THR
-70,-70,-90,-90,-90,-90,-90,-90,-90,-90,-90,-90,-90,-90,
-90,-90,-90,-90,-90,-90
SPUR:PEAK:THR?
>
-70.00,-70.00,-90.00,-90.00,-90.00,-90.00,-90.00,-90.00,
-90.00,-90.00,-90.00,-90.00,-90.00,-90.00,-90.00,-90.00,
-90.00,-90.00,-90.00,-90.00

Table of Contents