Chapter 2 SCPI Device Message Details
2-458
[:SENSe]:BPOWer|:TXPower:BURSt:WIDTh <time>
Burst Average Power Width
Function
This command sets the measurement width of the burst for Burst
Average Power measurement.
Command
[:SENSe]:BPOWer|:TXPower:BURSt:WIDTh <time>
Parameter
<time>
Measurement width (time)
Range Within the frequency range of trace display
Suffix code NS,US,MS,KS,S
S is used when omitted.
Example of Use
To set the burst width of Burst Average Power measurement to 20 ms.
BPOW:BURS:WIDT 20ms
[:SENSe]:BPOWer|:TXPower:BURSt:WIDTh?
Burst Average Power Width Query
Function
This command queries the measurement width of the burst for Burst
Average Power measurement.
Query
[:SENSe]:BPOWer|:TXPower:BURSt:WIDTh?
Response
<time>
Parameter
<time>
Measurement width (time)
Range Within the frequency range of trace display
Suffix code None. Value is returned in S units.
Example of Use
To query the burst width of Burst Average Power measurement.
BPOW:BURS:WIDT?
> 0.020000000